Image Problem: Color Variation.

  • Post author:
  • Post category:

Trouble Shooting tepCauseSolutionParts1Value wrong.Do the learning calibrration.Contact us.2Lower CIS defective.Replace the upper or lower CIS.AD6090 (V1)-Old: Upper & Lower CIS kit (003-8339-0-SP)AD6090 (V2)-New:Upper CIS (002-A344A-0-SP)Lower CIS (002-A357A-0-SP)3Main broad defective.Replace the…

Continue ReadingImage Problem: Color Variation.

Others: No Power.

  • Post author:
  • Post category:

Trouble Shooting StepCauseSolutionParts1Adapter defective.Replace the adapter.Adapter (005-3148-9-SP)2UI broad defective.Replace the UI board UIA216.UI board (004-2988-9-SP)3Main broad defective.Replace the main board MBA842.Main board AD6090 (004-3272-9-SP)AD6090N (004-3272A-9-SP)

Continue ReadingOthers: No Power.

Others: Rear Cover Opened.

  • Post author:
  • Post category:

Icon Trouble Shooting StepCauseSolutionParts1Rear cover open sensor defective.Replace the rear cover open sensor.Rear cover open sensor. (003-8219-0-SP)2Main broad defective.Replace the main board MBA842.Main board AD6090 (004-3272-9-SP)AD6090N (004-3272A-9-SP)

Continue ReadingOthers: Rear Cover Opened.

Error Code 2036: Ultrasonic Sensor Error.

  • Post author:
  • Post category:

Icon Trouble Shooting StepCauseSolutionParts1Ultrasonic receiver broad defective.Replace the ultrasonic receiver broad SBA277.Ultrasonic receiver broad (004-3201-9-SP)2Ultrasonic transmitter broad defective.Replace the ultrasonic transmitter broad SBA278.Ultrasonic transmitter board (004-3202-9-SP)3Main broad defective.Replace the main…

Continue ReadingError Code 2036: Ultrasonic Sensor Error.